Ozone Analyzer - Medium Range - UV-106M Rental

2b-Technology

Write a Review
California Regulatory Information

SKU: 012547
SKU:
012547
Condition:
New
Dimensions:
4" x 7.5" x 8.5"
Weight:
5 lbs.

Product Overview

2b Technology Gas Monitoring

The UV-106-M is a low range ozone analyzer featuring ozone measurements up to 1,000 ppm with low range measurement of ozone down to 0.01 ppm. Bench mount ozone analyzer is compact and easy to use.  The UV-106-M can be purchased in either the bench mount or NEMA enclosure. 

Integrated sample pump will pull a sample of ozone from your process or location into the analyzer and measure ozone in that gas stream.  A UV lamp and receptor are used to determine the quantity of ozone in gas by determining the rate of ozone absorption by the UV lamp.  This method of ozone detection will offer accurate detection of ozone at levels up to 1,000 without compromising low range detection below 1 ppm, while eliminating most cross-sensitivity from other gasses.

Features

  • Measurement based on UV absorption
  • Low power consumption; can be battery operated
  • Internal data logger with real time clock
  • On-board microprocessor with interactive menus includes data averaging options of 10 s, 1 m, 5 m and 1 hr
  • USB and RS-232 output of time/date, O3 concentration, internal temperature and pressure
  • Analog output (0-2.5 V and 4-20 mA) of ozone concentration in user selected units
  • Two level relay for control purposes (e.g., control of ozone source or turn warning light on and off)
  • Long-life pump (15,000 hr)
 
Optional Accessories Description
Intake Filter Kit PVDF Filter Housing 47 mm with 1/4" Fittings. Kit Includes: FHS (Filter Housing), and FPK-10 (Filters- Package of ten)
Portable B3 Battery Add portability to your UV-106 Analyzer with the B3 Battery! This battery allows you to take your UV-106 Analyzer anywhere.
CUV-1 Carrying Case This is an extremely rugged pelican case to carry your UV-106 in.

 

 More Information
 Weight 4.0000
 Electrical 11 - 28 VDC nominally 500 mA at 12V, 6.0 watt
 Dimensions 8.5" x 8.5" x 3.8" (21.6 x 21.6 x 9.5 cm)
 Warranty 1 Year
 Display Digital
 Detection Method UV Absorption
 Temperature Range 32 to 122°F (0.0 to 50°C)
 Accuracy Greater of 0.01 ppm or 2% of reading
 Resolution 0.01 ppm (10 ppb)
 Supporting Documents User manual and NIST traceable calibration sheet included.
 Internal Datalogging Yes, 32,736 Lines
 Display Units ppm
 Comms. Protocol RS-232, USB
 Zero Drift <0.01 ppm/day, <0.03 ppm/year
 LPM Flow Rate 1 LPM
 Relays 1
 Integrated Pump Yes
 Portability Yes, with purchase of B3 Battery
   
106M Manual
106M Manual
 
 

 

California Regulatory Warning

Example Image
Use in unoccupied spaces only.
Health hazard warning: emits ozone.

This product is not CARB Certified (Ozone Emissions for Air Cleaners).
Product Disclaimer: Does not meet California air cleaner regulation requirements; cannot be shipped to California.

Product Videos