Model 205 Ozone Analyzer

2b-Technology

Write a Review
California Regulatory Information


Pricing: Call for Price Today at 888.892.0303

SKU: 012483
SKU:
012483

Product Overview

Model 205 Dual Beam Ozone Monitor Overview

EPA Federal Equivalent Method (FEM)

The Model 205 Dual Beam Ozone Monitor makes use of two detection cells to improve precision, baseline stability, and response time. In the Dual Beam instrument, UV light intensity measurements Io (ozone-scrubbed air) and I (unscrubbed air) are made simultaneously. Combined with other improvements, this made it possible to reduce the time between ozone measurements to 2 seconds, making our instrument the fastest UV-based ozone monitor on the market, while still retaining the small size, weight, and power requirements of our popular Model 202 Ozone Monitor. Fast measurements are especially desirable for aircraft and balloon measurements where high spatial resolution is needed. For 10-s averaging, a precision of 1.0 ppb is achieved. Alternatively, one can choose to average more points internally to obtain even better precision. Flash card memory and a long-life, quiet internal air sampling pump are now standard on the instrument. The Model 205 Dual Beam Ozone Monitor has been approved by the EPA as a Federal Equivalent Method (FEM) based on a modification of the Model 202

Features

  • Quiet long-life internal air pump
  • Low power consumption (~8 watt)
  • Internal data logger (EEPROM), logs 16,383 lines of data
  • Real-time clock
  • Both serial and user-scalable analog voltage outputs (4-20 mA current output optional)
  • Flash card memory (SD card) for virtually unlimited, portable data storage
  • Three 0-2.5 VDC analog inputs for logging other measurements, such as external temperature, pressure, and humidity
  • Convenient user interface to microprocessor, including calibration parameters (gain and zero)
  • Selectable measurement time of 2 s or data averaging times of 10 s, 1 min, 5 min and 1 hr (custom averaging times available)
  • DewLine™ for elimination of any water vapor interference, a unique feature of 2B Tech instruments
  • Dual beam feature allows for heightened precision and improved baseline stability for applications where high spatial resolution is required.

Options

  • GPS for continuous logging of latitude, longitude, and altitude (for 10-s data mode and above)
  • Rechargeable lithium-ion battery provides 12 hours of continuous operation
  • External particle filter
  • Serial-to-USB adapter
  • 4-20 mA analog current output
  • Bluetooth for wireless data transmission
  • Low-temperature package for operation to -20 degrees C (pump, lamp heater)
  • High-altitude upgrade (pressure sensor for use up to 25 km altitude)
  • Extra long-life, low temperature internal air pump
  • Low-power option for even lower power consumption
  • Rack-mount case configuration

What Is Included?

  • Model 205 Dual Beam Ozone Monitor
  • Quiet, Long-Life Internal Air Pump
  • AC Power Adapter (100-240 VAC to 12 VDC) with Country-Specific Plug
  • Cigarette Lighter Adapter
  • Battery Wire Connector
  • Zeroing Cartridge
  • Flash Card Memory for virtually unlimited, portable data storage
  • SD Card (> 2 GB) and SD Card Reader
  • Operation Manual on USB Stick
  • Calibration Data and NIST-Traceable Calibration Certificate
  • Instrument Birth Certificate
  • One-Year Warranty
Specifications
Measurement Principle UV Absorption at 254 nm; Dual Beam
Federal Equivalent Method (FEM) Yes, EQOA-0410-190
Linear Dynamic Range 0 to 250 ppm
Resolution 0.1 ppb
Precision (1σ rms noise) Greater of 1.0 ppb or 2% of reading for 10-s average
Accuracy Greater of 1.0 ppb or 2% of reading
Limit of Detection (2σ) 2.0 ppb for 10-s average
NIST-Traceable Calibration Yes
Measurement Interval 2s (Data averaging options: 10 s, 1 min, 5 min, 1 hr)
Flow Rate (nominal) ~1.8 Liter/min
Flow Rate Requirement >1.2 L/min
Baseline Drift < 1 ppb/day
< 3 ppb/year
Sensitivity Drift < 1%/day
< 3%/year
Measurement Time, Frequency 2 s, 0.5 Hz
Response Time, 100% of Step Change 4 s, 2 points
Measurement and Averaging Times 2 s, 10 s, 1 min, 5 min, 1 hr
Ozone Units ppb, pphm, ppm, µg m-3, mg m-3
Pressure Units mbar, torr
Temperature Units °C, K
T and P Corrected Yes
Operating Temperature Range 0 to 50 °C standard; optional -20 to 50 °C with low temperature modifications (pump and lamp heater cold-weather package)
Operating Altitude Range ~0-13.5 km (150-1,013 mbar) standard; optional ~0-25 km (30-1,013 mbar) with upgraded pressure sensor
Power Requirement; Supplied by battery or 110/220 VAC Power Pack 12 V dc or 120/240 V ac, nominally 650 mA at 12 V, ~8 watt (~7 watt with cell heater unplugged)
Size 3.7 x 8.5 x 11.5 inches (9.5 x 21.5 x 30 cm)
Weight 5.8 lb (2.6 kg); 3.1 lb (1.4 kg) without instrument case
Internal Data Logger Capacity 16,383 lines (2-s fast measurement mode = 9.1 hr; 10 s avg. = 1.9 days; 1 min avg = 11 days; 5 min avg = 1.9 mo; 1 hr avg = 1.9 yr
SD Card Logger Capacity > 2 GB (> 5-year capacity for 10-s averaging mode)
Analog Inputs for Internal Logging of Other Instruments 3 Analog Inputs, 0-2.5 V; For example, could log external T, P, and RH
Data Outputs RS232, 0-2.5 V Analog, LCD Display, External USB Converter (Optional 4-20 mA Current; request quote)
Data Transfer Baud Rates 2400, 4800, 19200
Output Ranges User-Defined Scaling Factor in Menu
DewLine™ Yes
Quiet, Long-Life Internal Air Pump Yes
Flow Meter Yes
Options Battery; External Particle Filter; Serial-to-USB Adapter; GPS (for 10-s data mode and above); Bluetooth; 4-20 mA Current Output; Cold-Weather Package (pump, lamp heater); High-Altitude Upgrade (pressure sensor); Extra-Long-Life, Low-Temperature Internal Air Pump; Low-Power Option; Rack-Mount Case Configuration

Warranty Information

1 year warranty on defective parts, materials and workmanship.

California Regulatory Warning

Example Image
Use in unoccupied spaces only.
Health hazard warning: emits ozone.

This product is not CARB Certified (Ozone Emissions for Air Cleaners).
Product Disclaimer: Does not meet California air cleaner regulation requirements; cannot be shipped to California.